Metal & Susceptibility

Metal distorts the local magnetic field. Gradient echo suffers most (no refocusing). Higher bandwidth reduces distortion, shorter TE reduces dephasing, and higher field strength makes everything worse.

Acquiring…
APRL
Simulated image — educational phantom

Weighting

T2*-weighted gradient echo

Scan time

1:36

Voxel

0.90 × 1.20 × 5 mm

TE used

15 ms

SNR (vs reference protocol)100%
SAR (of normal-mode limit)0%
Chemical shift: 1.4 pxPhase lines acquired: 192Slices per TR: 26

PROTOCOL

Sequence

15 ms
160 Hz/px
5 mm

Field strength

MISSIONS

0/2 COMPLETE · IN ORDER
  1. Current mission: Shrink the artifact: spin-echo-based sequence and bandwidth ≥ 300 Hz/px+75 XP
  2. Locked: Make it as bad as possible: gradient echo, TE ≥ 30 ms, 3 T+50 XP