← SCANNER LAB
CHALLENGE MODE · 8 PROBLEMS
Fix My MRI
Each scan has a problem. Change the protocol until the goal is met. The simulated image updates as you go.
PROBLEM 01
Low SNR
The radiologist says the image is too grainy.
- SLICE
- BW
- NEX
- MATRIX
- +1
PROBLEM 02
Wrap-around
The back of the head is folding onto the front.
- FOV
- PHASE
- OVERSAMPLE
- MATRIX
PROBLEM 03
Respiratory ghosts
The patient can hold their breath for 20 seconds — but the sequence takes much longer.
- ETL
- R
- MATRIX
- NEX
- +2
PROBLEM 04
Chemical shift
Dark and bright bands around the kidneys are hiding the renal outline.
- BW
- PHASE
- FAT SAT
- NEX
- +1
PROBLEM 05
Too slow
The T2 takes over 4 minutes and the patient is in pain.
- ETL
- NEX
- MATRIX
- R
- +3
PROBLEM 06
SAR limit at 3 T
The sequence won't run: SAR exceeded.
- TR
- ETL
- SLICES
- TE
PROBLEM 07
FLAIR at 3 T
A FLAIR copied from 1.5 T doesn't null CSF on the new 3 T scanner.
- TI
- TR
- TE
PROBLEM 08
Metal artifact
A clip is blotting out the region of interest on gradient echo.
- SEQ
- TE
- BW
- SLICE